Original Research
Identifying HIV infection in South African women: How does a fourth generation HIV rapid test perform?
Submitted: 18 November 2011 | Published: 15 December 2011
About the author(s)
Kapila Bhowan, Paediatric HIV Diagnostic Syndicate, Wits Health Consortium, South AfricaEmma Kalk, Paediatric HIV Diagnostic Syndicate, Wits Health Consortium, South Africa
Sonjiha Khan, Paediatric HIV Diagnostic Syndicate, Wits Health Consortium, South Africa
Gayle Sherman, Paediatric HIV Diagnostic Syndicate, Wits Health Consortium, South Africa
Abstract
Objectives: We assessed the performance of the DC RT against third generation RT in antenatal and post-partum women.
Methods: Third generation RT Advance Quality and Acon were used in a serial algorithm to diagnose HIV infection in antenatal and post-partum women over six months at a tertiary hospital in Johannesburg, South Africa. This data provided the reference against which the DC RT was compared on plasma and whole blood samples.
Results: The 1019 participants comprised 345 (34%) antenatal and 674 (66%) post-partum women. Ninety women (8.8%) tested HIV-positive of whom 59 (66%) were tested antenatally, and 31 (34%) post-partum yielding prevalence rates of 17.1% and 4.6% respectively. The sensitivity and specificity of the Ab component of DC on plasma antenatally was 100% (93.8% – 100%) and 100% (98.6% – 100%) respectively and post-partum was 100% (88.9% – 100%) and 99.6% (98.8% – 99.9%) respectively. One false positive and not a single true positive p24Ag was detected. Of 505 post-partum women who tested HIV-negative 6–12 months prior to enrolment, 12 (2.4%) seroconverted.
Conclusion: The fourth generation DC offered no advantage over current third generation RT in the diagnosis of HIV infection.
Keywords
Metrics
Total abstract views: 10514Total article views: 23747
Crossref Citations
1. Improving the Quality of Diagnostic Studies Evaluating Point of Care Tests for Acute HIV Infections: Problems and Recommendations
Megan Smallwood, Nitika Pant Pai
Diagnostics vol: 7 issue: 1 first page: 13 year: 2017
doi: 10.3390/diagnostics7010013
2. Field accuracy of fourth-generation rapid diagnostic tests for acute HIV-1
Joseph M. Lewis, Peter Macpherson, Emily R. Adams, Eleanor Ochodo, Anita Sands, Miriam Taegtmeyer
AIDS vol: 29 issue: 18 first page: 2465 year: 2015
doi: 10.1097/QAD.0000000000000855
3. Evaluation of a Rapid Point of Care Test for Detecting Acute and Established HIV Infection, and Examining the Role of Study Quality on Diagnostic Accuracy: A Bayesian Meta-Analysis
Megan Smallwood, Rohit Vijh, Bénédicte Nauche, Bertrand Lebouché, Lawrence Joseph, Nitika Pant Pai, Robert L Schmidt
PLOS ONE vol: 11 issue: 2 first page: e0149592 year: 2016
doi: 10.1371/journal.pone.0149592
4. Poor Performance of the Determine HIV-1/2 Ag/Ab Combo Fourth-Generation Rapid Test for Detection of Acute Infections in a National Household Survey in Swaziland
Yen T. Duong, Yvonne Mavengere, Hetal Patel, Carole Moore, Julius Manjengwa, Dumile Sibandze, Christopher Rasberry, Charmaine Mlambo, Zhi Li, Lynda Emel, Naomi Bock, Jan Moore, Rejoice Nkambule, Jessica Justman, Jason Reed, George Bicego, Dennis L. Ellenberger, John N. Nkengasong, Bharat S. Parekh, A. M. Caliendo
Journal of Clinical Microbiology vol: 52 issue: 10 first page: 3743 year: 2014
doi: 10.1128/JCM.01989-14