Original Research

Assessment of the AQUIOS flow cytometer – An automated sample preparation system for CD4 lymphocyte PanLeucogating enumeration

Daniel Rhodes, Guislaine Carcelain, Mike Keeney, Christophe Parizot, Dominika Benjamins, Laurine Genesta, Jin Zhang, Justin Rohrbach, Denise Lawrie, Deborah K. Glencross
African Journal of Laboratory Medicine | Vol 8, No 1 | a804 | DOI: https://doi.org/10.4102/ajlm.v8i1.804 | © 2019 Daniel Rhodes, Guislaine Carcelain, Mike Keeney, Christophe Parizot, Dominika Benjamins, Laurine Genesta, Jin Zhang, Justin Rohrbach, Denise Lawrie, Deborah K. Glencross | This work is licensed under CC Attribution 4.0
Submitted: 13 March 2018 | Published: 05 December 2019

About the author(s)

Daniel Rhodes, Clinical Affairs, Beckman Coulter Immunotech, Marseille, France
Guislaine Carcelain, Immunology Laboratory, Assistance Publique Hopitaux De Paris, Paris, France
Mike Keeney, Lawson Health Research Institute, London Health Sciences Centre and St. Joseph’s Health Care, Victoria Hospital, London, Ontario, Canada
Christophe Parizot, Department of Immunology, University Hospital, Paris, France
Dominika Benjamins, London Health Sciences Centre, London, Ontario, Canada
Laurine Genesta, Biomnis Laboratory, Lyon, France
Jin Zhang, Life Science Flow Cytometry, Beckman Coulter Incorporated, Miami, Florida, United States
Justin Rohrbach, Clinical affairs, Beckman Coulter Incorporated, Miami, Florida, United States
Denise Lawrie, National Health Laboratory Service, Charlotte Maxeke Johannesburg Academic Hospital, Johannesburg, South Africa
Deborah K. Glencross, Faculty of Health Sciences, University of the Witwatersrand, Johannesburg, South Africa; and, National Health Laboratory Services, Johannesburg, South Africa

Abstract

Background: Flow cytometry has been the approach of choice for enumerating and documenting CD4-cell decline in HIV monitoring. Beckman Coulter has developed a single platform test for CD4+ T-cell lymphocyte count and percentage using PanLeucogating (PLG) technology on the automated AQUIOS flow cytometer (AQUIOS PLG).

Objectives: This study compared the performance of AQUIOS PLG with the Flowcare PLG method and performed a reference interval for comparison with those previously published.

Methods: The study was conducted between November 2014 and March 2015 at 5 different centres located in Canada; Paris, France; Lyon, France; the United States; and South Africa. Two-hundred and forty samples from HIV-positive adult and paediatric patients were used to compare the performances of AQUIOS PLG and Flowcare PLG on a FC500 flow cytometer (Flowcare PLG) in determining CD4+ absolute count and percentage. A reference interval was determined using 155 samples from healthy, non-HIV adults. Workflow was investigated testing 440 samples over 5 days.

Results: Mean absolute and relative count bias between AQUIOS PLG and Flowcare PLG was −41 cells/µL and −7.8%. Upward and downward misclassification at various CD4 thresholds was ≤ 2.4% and ≤ 11.1%. The 95% reference interval (2.5th – 97.5th) for the CD4+ count was 453–1534 cells/µL and the percentage was 30.5% – 63.4%. The workflow showed an average number of HIV samples tested as 17.5 per hour or 122.5 per 8-hour shift for one technician, including passing quality controls.

Conclusion: The AQUIOS PLG merges desirable aspects from conventional flow cytometer systems (high throughput, precision and accuracy, external quality assessment compatibility) with low technical operating skill requirements for automated, single platform systems.


Keywords

HIV; cluster of differentiation 4, CD4 Enumeration; PanLeucogating

Metrics

Total abstract views: 5947
Total article views: 8433

 

Crossref Citations

1. Cross‐sectional reassessment after 4 years of clinical routine use of AQUIOS CL for absolute T cell quantitation in a university hospital
Léa Lemoine, Christophe Malcus, Françoise Poitevin‐Later, Fabienne Venet, Guillaume Monneret, Morgane Gossez
Cytometry Part B: Clinical Cytometry  vol: 102  issue: 5  first page: 390  year: 2022  
doi: 10.1002/cyto.b.22069

2. Newly implemented community CD4 service in Tshwaragano, Northern Cape province, South Africa, positively impacts result turn-around time
Lindi-Marie Coetzee, Naseem Cassim, Deborah K. Glencross
African Journal of Laboratory Medicine  vol: 11  issue: 1  year: 2022  
doi: 10.4102/ajlm.v11i1.1376

3. Metrological traceability in flow cytometry? Evaluation of a new volumetric method for lymphocyte subsets
Rosalie Y. Ward, Matthew Stevens, Sheriaz Bashir
International Journal of Laboratory Hematology  vol: 46  issue: 3  first page: 488  year: 2024  
doi: 10.1111/ijlh.14219