Abstract
Background: Understanding factors that impact HIV viral load (VL) accuracy in resource-limited settings is key to quality improvement.
Objective: We evaluated whether testing delay and specimen storage between 25 °C and 30 °C before testing affected results.
Methods: Between November 2019 and June 2023, 249 individuals on antiretroviral therapy, or with newly diagnosed HIV, were recruited from clinics in Cape Town and Gqeberha, South Africa, and three plasma preparation tubes were collected. One tube was tested within 24 h, while the others were stored uncentrifuged at ambient temperatures before testing. Centrifugation and testing of matched samples were performed on Day 4 and Day 7 after collection.
Results: Time delay and ambient storage had minimal impact in specimens with a Day 1 VL of > 100 copies/mL. When grouped by Day 1 VL range, 96% – 100% of specimens at Day 4 and 93% – 100% at Day 7 had VLs within 0.5 log copies/mL of the first result. The greatest variability at Days 4 and 7 was observed when the Day 1 VL was < 100 copies/mL. However, there was no trend of increasing difference over time. Of Day 1 specimens with undetectable VL, or VL < 50 copies/mL, 80% had concordant results at Day 4 and 78% at Day 7.
Conclusion: These results show that VL is stable in plasma preparation tubes for 7 days when stored at room temperature. There is significant variability in specimens with low VL, but variability is not affected by testing delay.
What this study adds: Ideal HIV VL testing conditions are frequently unachievable in resource-limited settings. Data are needed on whether this impacts on the validity of test results. Our results provide reassurance that storage at ambient temperature for up to 7 days before testing does not substantially affect the VL result.
Keywords: HIV viral load stability; delayed testing; plasma preparation tubes; high-throughput viral load testing; diagnostic accuracy.
Introduction
Viral load (VL) testing is an essential tool for monitoring patients on anti-retroviral therapy (ART) and, at a programmatic level, provides data to assess the Joint United Nations Programme on HIV/AIDS 95:95:95 targets,1 specifically the third target which defines the proportion of patients on ART that are virologically suppressed. South Africa has the largest ART treatment programme in the world, with > 6 million people on therapy.2 In 2022, more than 6 m VL tests were performed as part of this programme.3 South Africa’s VL testing model involves centralised testing in high-throughput laboratories. Major challenges that have the potential to compromise test accuracy relate to logistics and pre-analytical problems, which include sub-optimal specimen collection, transport, storage, handling, and testing delay.
Currently, VL platform manufacturers advise that plasma should be separated within 6 h to 24 h of collection, then stored at 4 °C or frozen at −80 °C if not tested within 5 days.4,5 In resource-constrained environments, specimens frequently reach testing sites beyond the recommended time after collection and may be subjected to further delays before testing takes place. The extent to which these pre-analytical factors compromise the accuracy of results has been evaluated in laboratory simulation studies,6,7 and also a systematic review.8 Findings suggest that viral RNA is preserved beyond the currently recommended testing time. Good preservation of viral RNA was also found in a local study performed at Groote Schuur Hospital (GSH) using routine diagnostic specimens collected in ethylenediaminetetraacetic acid tubes or ethylenediaminetetraacetic acid-plasma preparation tubes (PPT). In this study, diagnostic specimens were stored after initial HIV VL testing at a range of times and storage temperatures and then re-tested.9 The VL in specimens stored for up to 1 week had very good concordance with the first result.9 However, tubes had already been centrifuged prior to storage and the study was performed entirely in the laboratory environment. In addition, it was only possible to follow trends in VL from the same patient over 2 time points due to specimen volume constraints. More studies are therefore needed to gain a better understanding of the factors that affect VL test reliability, as this knowledge is critical for improving the standard of testing in resource constrained environments, like South Africa. To expand on our previous findings, a field study was undertaken where participants were recruited to provide additional specimens for VL testing. By means of serial testing of specimens from the same participants we evaluated the extent to which testing delay and adverse storage (including storage temperature and centrifugation) affects the reliability of VL results.
Methods
Ethical considerations
Ethical approval for the study was obtained from the Human Subjects Research Ethics Committee of the Faculty of Health sciences at the University of Cape Town (HREC Ref 159/2019). Written informed consent was obtained from all participants prior to enrolment. Only individuals of 18 years or older were recruited. Study-related data were stored on a password-protected device to which only the principal investigator had access.
Study objectives
The study objectives were as follows: to evaluate the impact of delayed testing on the reliability of HIV RNA quantification in diagnostic specimens, collected and stored in un-centrifuged PPT tubes and to describe the impact of storage at a warmer ambient temperature (storage between 25 °C and 30 °C) in diagnostic specimens in un-centrifuged PPT tubes. We also evaluated whether centrifugation (enabling physical separation of plasma from cells) prior to storage improves the concordance of results where testing is delayed (evaluated at one site, namely GSH).
Study design
The study was performed at three sites within the National Health Laboratory Service network of HIV VL testing laboratories. Patients were recruited from four ART clinics, three in the Cape Metro and one in the Eastern Cape Province of South Africa. Enrolment for two sites (Gqeberha/Port Elizabeth [PE] and Tygerberg Hospital [TBH]) ran from 21 November 2019 to 24 January 2020 for PE, and 05 December 2019 to 23 January 2020 for TBH. Enrolment for the third site (GSH) began on 28 August 2022 and ended on 06 June 2023 (following normalisation of laboratory services after the severe acute respiratory virus-coronavirus 2 pandemic).
Patients who were due to have a VL test or who were newly diagnosed with HIV and not yet on therapy, were approached, and consented to give two extra blood tubes (5 mL – 7 mL per tube) for HIV VL testing. Agnostic patient recruitment was not done as we were not trying to collect a sample set reflective of the HIV-positive population in South Africa, but rather wanted to evaluate a technical issue in stored specimens. Thus, we aimed to collect roughly half of the specimens from patients with low-level or undetectable viral loads (on ART) and half from patients who were expected to have unsuppressed VLs (not yet started on antiretroviral drugs).
Three PPT tubes were collected during the same blood draw and delivered to the regional HIV VL testing laboratory (Figure 1). On arrival, one tube was centrifuged (at approximately 3000 g for 10 min) and sent for immediate HIV VL testing. Testing was completed within 24 h of sample collection. The result from this test was issued for routine patient management. The other two tubes were stored uncentrifuged in routine field conditions where the temperature varied between 25 °C and 30 °C. Four days after collection, one of the stored tubes was centrifuged (at 3000 g for 10 min) and a VL test was performed. Similarly, 7 days after collection, the third tube was centrifuged and a VL test was performed. Viral load testing was performed using the routine assay in each of the three participating laboratories at the time of study, namely Roche CAP/CTM (TBH laboratory), Roche 6800 (PE laboratory) (both Roche Molecular Systems, Pleasanton, California, United States) and Abbott Alinity m (GSH laboratory; Abbott Laboratories, Chicago, Illinois, United States). On a subset of patients (103 specimens, done at GSH laboratory only), HIV VL was repeated at Day 7 on the tube that had been centrifuged and tested on Day 1. The Day 1 centrifuged tubes were stored under the same conditions as the yet untested ones, prior to repeat testing. The tubes were not re-centrifuged prior to repeat testing on Day 7. They were used to assess whether storage in an already centrifuged state improved stability of VL in the sample.
|
FIGURE 1: Study plan for testing factors affecting accuracy of viral load samples, South Africa, 2019–2023. |
|
Data analysis
Data analysis was performed using Microsoft Excel (Microsoft Corporation, Redmond, Washington, United States). Bland-Altman plots were used to evaluate differences in Day 1 versus Day 4 VL results and between Day 1 and Day 7.10 Viral load values were log transformed as this is the standard way of expressing these values in clinical practice.
For analysis, unless otherwise specified, samples with a VL that was lower than the limit of detection (LDL) were assigned a value of ‘1’ (0 log) copies/mL. Results with a VL value of < 20 copies/mL were assigned a value of ‘19’ (1.28 log) copies/mL, otherwise the log value of the reading in copies/mL was used for analysis. We considered a difference in VL of < 0.3 log (or 2-fold difference) to reflect ‘no change’ as this is within the expected variability of this technology. A change of > 0.5 log copies/mL (> 3-fold difference) was considered likely to reflect a change that was caused by factors other than random expected variation.11 In addition, the results were evaluated in terms of whether the results obtained at Day 4 or Day 7 resulted in a reclassification of a patients’ VL control status. For this analysis: a viral load of ≤ 0 copies/mL was considered to reflect viral suppression;12,13 a viral load of 50 copies/mL–1000 copies/mL was considered to reflect low-level viraemia (LLV);12,13 and a VL value of > 1000 copies/mL was considered to reflect unsuppressed VL or viral failure.12,13
Results
A total of 249 participants were consented and provided 3 PPT tubes for testing (51 from TBH, 60 from Gqeberha (PE) and 138 from GSH. Details from each study site are given in Table 1. Specimens from 6 patients arrived at the laboratory more than 24 h after collection and these were excluded from the study. Altogether, 123 participants were already on ART and 120 were newly diagnosed as HIV-positive, but not yet on ART.
TABLE 1: Characteristics of patients and testing sites used for viral load test accuracy study, South Africa, 2019–2023. |
Day 1 VLs were segregated into the following VL categories: 55 (22.6%) were LDL; 46 (18.9%) were detectable, but < 50 copies/mL; 7 (2.9%) were 50 copies/mL – 100 copies/mL; 28 (11.5%) were 2–3 log copies/mL; 30 (12.3%) were 3–4 log copies/mL; 36 (14.8%) were 4–5 log copies/mL; 30 (12.3%) were 5–6 log copies/mL; and 11 (4.5%) were > 6 log copies/mL (Table 2).
TABLE 2: Proportion of samples tested on Day 4 and Day 7 within 0.3 log copies/mL and 0.5 log copies/mL of Day 1 viral load value, by Day 1 VL category, South Africa, 2019–2023. |
Applying current guidelines to define the participants’ HIV control status, as determined by the Day 1 VL value, 101 participants would have been classified as virally suppressed, 35 had LLV and 107 had unsuppressed VLs (Table 3). For reference, the full data set containing the matched Day 1, Day 4, and Day 7 VL results for each patient is available as Online Supplementary Document 1.
TABLE 3: Viral load concordance on Day 4 and Day 7 according to viral control status as determined by the Day 1 test, South Africa, 2019–2023. |
Viral load concordance at Days 1, 4 and 7
Overall, there was very little difference in paired readings at both Day 4 and Day 7 compared with Day 1. The mean bias at Day 4 was 0.15 and for Day 7 was 0.12 log copies per mL, indicating a higher VL reading in the samples where testing was delayed (Figure 2). For samples with an initial VL of > 2 log copies/mL, the paired VL value at Day 4 and Day 7 was within 0.5 log copies/mL in 96% – 100% of instances for Day 4 and 93% – 100% of instances for Day 7 (Table 2), signifying minimal impact of the testing delay on the quantification. The greatest variation was observed for specimens with Day 1 VLs less than 2 log copies/mL.
|
FIGURE 2: Variability of all paired viral load results at Days 4 and 7, South Africa, 2019–2023. Bland-Altman plots of Day 1 versus Day 4 VL (a) and Day 1 versus Day 7 (b) show that viral load measurements tended to be systematically higher by an average of 0.12 log at Day 4 and 0.14 log at Day 7 as compared to Day 1. The bias tended to be greater in samples with low RNA levels (< 2 log copies/mL); (c) shows variability of paired viral loads of Day 1 samples that were re-run (R) on Day 7, with the Bland Altman plot revealing minimal bias and even lower variance in Day 1 versus Day 7 results. Green lines represent +2 and −2 standard deviations and the red dotted line represents the mean difference in HIV viral load on the Bland Altman plots. |
|
When differences in paired VLs were evaluated using important thresholds that define clinical response, namely viral suppression (VL undetectable or < 50 copies/mL), LLV (VL 50 copies/mL – 1000 copies/mL) and viral failure/unsuppressed VL (> 1000 copies/mL) (Table 3), similarly good concordance was observed, with most specimens at Days 4 and 7 having VL in the same clinical range as Day 1. Lowest agreement was in the 101 samples with a Day 1 VL below 50 copies/mL, but nonetheless 81% remained < 50 copies/mL at the Day 4 test and 79% at the Day 7 test. Most of the discordant results, 17 specimens at Day 4 and 19 specimens at Day 7, shifted into the LLV category, while for 2 specimens at Day 4 and 2 specimens at Day 7, the VL was > 1000 copies/mL.
Concordance was higher, at 86% in specimens that were LLV on Day 1 testing at both Days 4 and 7, and even higher, at 96% in specimens where the Day 1 VL was > 1000 copies/mL.
Patients with a Day 1 VL that was detectable, but < 50 copies/mL, were 2.6 times more likely to have a detectable VL at Day 4 and 3.4 times more likely to have a detectable VL at Day 7 than patients with an undetectable Day 1 VL (LDL).
Overall, detection in a later specimen appeared to be stochastic. There was no trend to suggest that this phenomenon worsened with time. Its frequency appeared to be affected by testing site and technology used, as it varied from 0% (GSH, Abbott Alinity), 11% and 22% at Days 4 and 7 (TBH, Roche CAP/CTM) to 50% (PE, Roche Cobas 6800).
Viral load concordance in samples centrifuged on Day 1
Storage of specimens already centrifuged appeared to improve concordance. The Day 7 VL values in specimens that were stored in a centrifuged condition (plasma separated from the cells) were very similar to the Day 1 value (Figure 2c). The mean difference in VL between Day 1 and Day 7 was 0.03 log copies/mL and standard deviation was narrower, at 0.42, than for the specimens that were first centrifuged and tested on Day 7. Only one result had a clinically significant difference (–1.1 log copies/mL) in VL between Day 1 (4.95 log copies/mL) and Day 7 (3.85 log copies/mL).
Discussion
In this field study conducted in Cape Town and Gqeberha in South Africa between 2019 to 2023, we found that delay of HIV VL testing for up to 7 days after specimen collection and storage at ambient temperature had minimal effect on test results. These findings are compatible with previous work done in South Africa in 2017, using stored laboratory specimens,9 and also support the findings of a systematic review by Bonner et al.8 The study conditions were designed to match more closely the real-world situation in low- and middle-income countries with high HIV disease burdens and large ART programmes, where testing is often delayed and storage temperatures are higher than recommended due to logistical issues. These results provide reassurance that time delay and ambient storage is not a major cause of result inaccuracies.
Most of the variability that resulted in misclassification of a patient’s status was observed in specimens with Day 1 VLs of < 2 log copies/mL (from virally suppressed patients and those with LLV). In particular, a proportion of results in the viral suppression range shifted into the LLV range at one or both later time points, and some that were LLV shifted into the viral suppression range. The most probable explanation for this is the inherent variability of real-time polymerase chain reaction technology in specimens with target levels close to the limit of detection.14 In such specimens, target may be only intermittently detected in test replicates, and greater variability can be expected in the quantified value.15 Of note is that specimens from patients that were virally suppressed, but had a detectable VL that was < 50 copies/mL at Day 1, were 2 times to 3 times more likely to have a detectable VL in a stored specimen than if the VL was undetectable at Day 1. Low-level viraemia reflects a state of incomplete viral suppression and has been associated with increased risk of VL failure in the future.16,17,18 Because of this, there is a move to set the VL indicative of viral failure (currently defined as > 1000 copies/mL) at a lower VL threshold, such as at 200 or 50 copies/mL. As can be seen in this study, this can be expected to increase the chance of misclassification of patients due to the inherent variability of real-time polymerase chain reaction in this range. Confirmation of viraemia on a second specimen would help to mitigate this and is recommended in current guidelines.12,13
Occasional specimens displayed a higher-than-expected variability, namely 4 that were virally suppressed on Day 1, but > 1000 copies/mL at either Day 4 or Day 7. It is probable that factors other than time delay were responsible. Potential explanations for this stochastic variability could be specimen contamination with HIV RNA during processing, inadequate centrifugation prior to testing or some other undefined laboratory factor.19,20,21 Technical factors clearly played a role in this study, as the proportion of specimens that were LDL on Day 1, but detectable later, varied markedly at the different testing sites.
One additional measure to improve test concordance that was evaluated was centrifugation of specimens prior to storage. In a subset of specimens, the same tube that was centrifuged and tested on Day 1, was rerun on Day 7. These results showed very low variability and no specimen had a clinically actionable change in VL when comparing the Day 1 and Day 7 results. A caveat is that this experiment was only done at the site where there was very low variability in the stored specimens anyway.
Limitations
Only three specimens were collected from each patient, and this limited the number of factors we could evaluate. Testing occurred across three sites and different testing platforms could have accounted for some of the differences that were seen, for example a much lower rate of detectable viral RNA in later specimens with Day 1 suppressed VL at one site. These will need to be evaluated at a later stage. Specimens were only tested once on Day 1. A repeat test on Day 1 would have provided key data on the inherent variability of VL results on our platforms and given context to the variability observed at the later time points.
Conclusion
Our field study provides further evidence that time delay has minimal impact on VL quantification when specimens are stored at room temperature for up to 7 days. This extends the time allowable before VL testing needs to be done. Most of the discordant results were in the LLV range and likely reflect inconsistent detection at the limit of detection of the assay. Centrifugation of specimens prior to storage improves the concordance of results at later time points and early centrifugation after collection should be prioritised for HIV testing programmes. The contribution of other patient-related and technical issues requires further investigation.
Acknowledgements
We thank Widaad Zemanay and Nchimunya Hapepa for help with project management and Sr Dyonase for assistance with patient recruitment. We also thank staff at National Health Laboratory Service laboratories with help processing study samples.
Competing interests
The authors declare that they have no financial or personal relationships that may have inappropriately influenced them in writing this article.
Authors’ contributions
D.R.H., S.K., N.-Y.H., H.N., L.H. and G.v.Z. all contributed to designing the study. H.N. recruited patients and oversaw testing at the Port Elizabeth (Gqeberha) laboratory. J.R. recruited patients and oversaw testing at the Tygerberg laboratory. J.O. recruited patients and D.R.H. oversaw testing at the Groote Schuur Hospital laboratory. D.R.H. wrote the manuscript and all authors contributed to data analysis and reviewing of the final manuscript.
Sources of support
This research received no specific grant from any funding agency. The National Health Laboratory Service provided support for testing of study samples. No other formal funding was received for this study.
Data availability
The authors confirm that the data supporting the findings of this study are available within its supplementary materials.
Disclaimer
The views and opinions expressed in this article are those of the authors and are the product of professional research. It does not necessarily reflect the official policy or position of any affiliated institution, funder, agency, or that of the publisher. The authors are responsible for this article’s results, findings, and content.
References
- Robert B, Brown EB. FAST-TRACK ending the AIDS epidemic by 2030. UNAIDS Publ [serial online]. 2014;(1):1–14. [cited 2024 Apr 01]. Available from: https://www.unaids.org/sites/default/files/media_asset/JC2686_WAD2014report_en.pdf
- NHLS Annual report 2017/18 [homepage on the Internet]. 2018, p. 71–75. [cited 2024 Apr 01]. Available from: http://intranet.nhls.ac.za/assets/files/policy/NHLS_AR_2018.pdf
- Hans L, Cassim N, Sarang S, et al. HIV viral load testing in the South African public health setting in the context of evolving ART guidelines and advances in technology, 2013–2022. Diagnostics. 2023;13(17):2013–2022. https://doi.org/10.3390/diagnostics13172731
- Abbott RealTime HIV1 kit insert 51-602100/R10 [homepage on the Internet]. 2014 [cited 2024 Apr 01]. Available from: https://www.molecular.abbott/content/dam/add/molecular/products/pdf-/realtime-hiv-1-package-insert.pdf
- Roche CAP/CTM HIV-1 v2.0 EXPT-IVD [homepage on the Internet]. 2018 [cited 2024 Apr 01]. Available from: https://elabdoc-prod.roche.com/eLD/api/downloads/ee10c6ed-0bff-e911-fa90-005056a71a5d?countryIsoCode=za
- Vandamme AM, Van Laethem K, Schmit JC, et al. Long-term stability of human immunodeficiency virus viral load and infectivity in whole blood. Eur J Clin Invest. 1999;29(5):445–452. https://doi.org/10.1046/j.1365-2362.1999.00462.x
- Amellal B, Murphy R, Maiga A, et al. Stability of HIV RNA in plasma specimens stored at different temperatures. HIV Med. 2008;9(9):790–793. https://doi.org/10.1111/j.1468-1293.2008.00632.x
- Bonner K, Siemieniuk RA, Boozary A, Roberts T, Fajardo E, Cohn J. Expanding access to HIV viral load testing: A systematic review of RNA stability in EDTA tubes and PPT beyond current time and temperature thresholds. PLoS One. 2014;9(12):1–13. https://doi.org/10.1371/journal.pone.0113813
- Hardie D, Korsman S, Ameer S, Vojnov L, Hsiao NY. Reliability of plasma HIV viral load testing beyond 24 hours: Insights gained from a study in a routine diagnostic laboratory. PLoS One. 2019;14(7):1–12. https://doi.org/10.1371/journal.pone.0219381
- Giavarina D. Understanding Bland Altman analysis. Biochem Med. 2015;25(2):141–151. https://doi.org/10.11613/BM.2015.015
- Wilson D, Naidoo S, Bekker L-G, Mark Cotton GM, editors. Handbook of HIV medicine. 1st ed. Oxford: Oxford University Press, 2002; p. 44–46.
- World Health Organization. Consolidated guidelines on HIV prevention, testing, delivery and monitoring: Treatment, service recommendations for a public health approach. Geneva: World Health Organization; 2021.
- Nel J, Meintjes G, Osih R, et al. Southern African HIV Clinicians Society guidelines for antiretroviral therapy in adults [homepage on the Internet]. 2023; p. 1–78. [cited 2024 Apr 01]. Available from: http://www.sahivsoc.org
- Giordano M, Kelleher T, Colonno RJ, Lazzarin A, Squires K. The effects of the Roche AMPLICOR HIV-1 MONITOR® UltraSensitive Test versions 1.0 and 1.5 viral load assays and plasma collection tube type on determination of response to antiretroviral therapy and the inappropriateness of cross-study comparisons. J Clin Virol. 2006;35(4):420–425. https://doi.org/10.1016/j.jcv.2005.10.011
- Newman H, Maritz J. Basic overview of method validation in the clinical virology laboratory. Rev Med Virol. 2017;27(5):1–7. https://doi.org/10.1002/rmv.1940
- Chun HM, Abutu A, Milligan K, et al. Low-level viraemia among people living with HIV in Nigeria: A retrospective longitudinal cohort study. Lancet Glob Health. 2022;10(12):e1815–e1824. https://doi.org/10.1016/S2214-109X(22)00413-2
- Crespo-Bermejo C, de Arellano ER, Lara-Aguilar V, et al. Persistent low-level viremia in persons living with HIV undertreatment: An unresolved status. Virulence. 2021;12(1):2919–2931. https://doi.org/10.1080/21505594.2021.2004743
- Ellman TM, Alemayehu B, Abrams EJ, Arpadi S, Howard AA, El-Sadr WM. Selecting a viral load threshold for routine monitoring in resource-limited settings: Optimizing individual health and population impact. J Int AIDS Soc. 2017;20:16–18. https://doi.org/10.1002/jia2.25007
- Procop GW, Taege AJ, Starkey C, et al. Preanalytic process linked to spuriously elevated HIV viral loads: Improvement on an FDA-approved process. Diagn Microbiol Infect Dis. 2017;89(1):44–46. https://doi.org/10.1016/j.diagmicrobio.2016.09.003
- Kran AMB, Jonassen TØ, Sannes M, et al. Overestimation of human immunodeficiency virus type 1 load caused by the presence of cells in plasma from plasma preparation tubes. J Clin Microbiol. 2009;47(7):2170–2174. https://doi.org/10.1128/JCM.00519-09
- Salimnia H, Moore EC, Crane LR, MacArthur RD, Fairfax MR. Discordance between viral loads determined by Roche COBAS AMPLICOR human immunodeficiency virus type 1 monitor (version 1.5) standard and ultrasensitive assays caused by freezing patient plasma in centrifuged Becton-Dickinson vacutainer brand plasma prepa. J Clin Microbiol. 2005;43(9):4635–4639. https://doi.org/10.1128/JCM.43.9.4635-4639.2005
|